光容积图
夹紧
血压
血压计
袖口
脉动流
医学
波形
生物医学工程
持续监测
计算机科学
心脏病学
工程类
内科学
外科
计算机视觉
运营管理
雷达
滤波器(信号处理)
电信
作者
Ramakrishna Mukkamala,Jin‐Oh Hahn,Anand Chandrasekhar
出处
期刊:Elsevier eBooks
[Elsevier]
日期:2021-11-12
卷期号:: 359-400
被引量:13
标识
DOI:10.1016/b978-0-12-823374-0.00010-4
摘要
Clinical management of hypertension and hypotension has been hindered by current cuff blood pressure (BP) measurement devices, which are not always readily available or continuous and can be disruptive to the user. Photoplethysmography (PPG) is a simple yet effective tool for tracking pulsatile arterial blood volume changes that can improve BP measurement. Volume clamping via finger cuff-PPG devices is a proven continuous method, whereas oscillometry via PPG-force sensor units, pulse transit time detected via PPG waveform(s), and PPG waveform feature extraction are potential cuff-less methods. The objective of this chapter is to facilitate the advancement of PPG-based BP monitoring. First, the clinical BP measurement methods and their limitations are explained. Then, the PPG principle and available sensors are reviewed. Next, the four PPG-based BP measurement methods are described including principles, embodiments, and empirical evidence. Finally, these methods are compared, future research directions are pinpointed, and a brief outlook on the approach is provided.
科研通智能强力驱动
Strongly Powered by AbleSci AI