Christel H. Kamani,Wenjie Huang,Joel Lutz,Andreas A. Giannopoulos,Dimitri Patriki,Elia von Felten,Moritz Schwyzer,Catherine Gebhard,Dominik C. Benz,Tobias A. Fuchs,Christoph Gräni,Aju P. Pazhenkottil,Philipp A. Kaufmann,Ronny R. Buechel
标识
DOI:10.1016/j.acra.2020.10.025
摘要
To evaluate the impact of adaptive statistical iterative reconstruction-V (ASIR-V) on the accuracy of ultra-low-dose coronary artery calcium (CAC) scoring.One-hundred-and-three patients who underwent computed tomography (CT) for CAC scoring were prospectively included. All underwent standard scanning with 120-kilovolt-peak (kVp) and with 80- and 70-kVp tube voltage. ASiR-V was applied to the 80- and 70-kVp scans at different levels. The 120-kVp scans reconstructed with filtered back projection served as the standard of reference. Recently published novel kVp-adapted thresholds were used for calculation of CAC scores from 80- and 70-kVp scans and the resulting CAC scores were compared against the standard of reference. Patients were stratified into six CAC score risk categories: 0, 1-10, 11-100, 101-400, 401-1000, and >1000.Increasing levels of ASIR-V led to an increasing underestimation of CAC scores with bias ranging from -128 to -118 and from -205 to -198 for the 80- and 70-kVp scans, respectively, when compared with the standard of reference. Reconstruction with 20% and 40% ASIR-V for the 80- and 70-kVp scans, respectively, yielded noise levels comparable to the standard of reference. Nevertheless, a change in risk-class was observed in 29 (28.6%) and 46 (44.7%) patients, exclusively to a lower risk-class, when CAC scores were derived from these reconstructions.ASIR-V leads to noise reduction in CT scans acquired with low tube-voltages. However, ASIR-V introduces substantial inaccuracies and marked underestimation of ultra-low-dose CAC scoring as compared with standard-dose CAC scoring despite normalization of noise.
科研通智能强力驱动
Strongly Powered by AbleSci AI