Iodine is essential for human growth and can enter the body through food, water, and air. Analyzing its presence in the environment is crucial for ensuring healthy human development. However, current large-scale instruments have limitations in the field analysis of iodine. Herein, a miniaturized purge and trap point discharge microplasma optical emission spectrometric (P&T-μPD-OES) device was developed for the field analysis of iodine in water. Volatile iodine molecules were produced from total inorganic iodine (TII) through a basic redox reaction under acidic conditions, then the purge and trap module effectively separated and preconcentrated iodine molecules. The iodine molecules were subsequently atomized and excited by the integrated point discharge microplasma and an iodine atomic emission line at 206.24 nm was monitored by the spectrometer. Under optimal conditions, this proposed method had a detection limit of 16.2 μg L