生物传感器
材料科学
聚二甲基硅氧烷
重复性
膜
纳米技术
胶体金
表面改性
色谱法
纳米颗粒
化学工程
化学
生物化学
工程类
作者
Qiang Zhang,Dong Zhao,Xiushan Dong,Qianqian Duan,Jianlong Ji,Yan Liu,Zhen Pei,Chao Ji,Shengbo Sang
标识
DOI:10.1002/admi.202101461
摘要
Abstract The membrane based on receptor functionalization provides a new paradigm for the development of mechanical biosensors. However, improvement of sensitivity and test accuracy is still a challenge for mechanical biosensors in practical application. Herein, a surface stress mechanical biosensor (MBioS) based on double‐side‐gold nanoparticale (AuNP)‐coated grid‐type polydimethylsiloxane (PDMS) membrane (D‐G‐MBioS) and 3D printing for human serum albumin (HSA) detection is developed. The surface stress is amplified by the grid coupling sandwich immune structure to improve the sensitivity of the MbioS, successfully reducing limite of detection (LOD) by two orders of magnitude. By self‐assembly of AuNPs, the double‐side‐coated PDMS membrane is prepared, and the detection accuracy of the sensor is improved. The 3D printing is employed to improve the consistency of the MBioS. The D‐G‐MBioS system is constructed, realizing the HSA detection in range of 0–50 µg mL −1 with a LOD of 20.6 ng mL −1 . In addition, the D‐G‐MBioS based on sandwich assay is exhibited the high specificity, selectivity, stability, and repeatability. Especially, the D‐G‐MBioS can detect the HSA in real samples. This study provides a way to improve the sensitivity of mechanical biosensor.