Paulo E. Cruvinel,S. Mascarenhas,J. Miranda,Robert G. Flocchini
出处
期刊:IEEE Transactions on Nuclear Science [Institute of Electrical and Electronics Engineers] 日期:1992-02-01卷期号:39 (1): 25-28被引量:7
标识
DOI:10.1109/23.120132
摘要
Using a Perovskite crystal, a thermal detector has been designed for measurements of proton beam currents. For a given energy, the detector has a linear response with current intensity and an inverse response with chopping frequency. Measurements up to 50 nA (4.5 MeV H/sup +/) were carried out using a cyclotron, and a calibration curve is presented. The detector may be used over a wide range of energies, has a low cost, and is simple to construct. In addition, it can be used inside or outside vacuum, and it does not require an external bias field.< >