超声波
医学
分割
股神经
阔筋膜
三维超声
股动脉
深筋膜
解剖
生物医学工程
计算机科学
放射科
人工智能
作者
Erik Smistad,Daniel Høyer Iversen,Linda Leidig,Janne Beate Lervik Bakeng,Kaj Fredrik Johansen,Frank Lindseth
标识
DOI:10.1016/j.ultrasmedbio.2016.08.036
摘要
Ultrasound-guided regional anesthesia can be challenging, especially for inexperienced physicians. The goal of the proposed methods is to create a system that can assist a user in performing ultrasound-guided femoral nerve blocks. The system indicates in which direction the user should move the ultrasound probe to investigate the region of interest and to reach the target site for needle insertion. Additionally, the system provides automatic real-time segmentation of the femoral artery, the femoral nerve and the two layers fascia lata and fascia iliaca. This aids in interpretation of the 2-D ultrasound images and the surrounding anatomy in 3-D. The system was evaluated on 24 ultrasound acquisitions of both legs from six subjects. The estimated target site for needle insertion and the segmentations were compared with those of an expert anesthesiologist. Average target distance was 8.5 mm with a standard deviation of 2.5 mm. The mean absolute differences of the femoral nerve and the fascia segmentations were about 1–3 mm.
科研通智能强力驱动
Strongly Powered by AbleSci AI