Martin Malina,Bengt Lindblad,Krasnodar Ivancev,Mats Lindh,Janne Malina,Jan Brunkwall
标识
DOI:10.1177/152660289800500404
摘要
To investigate if stents with hooks and barbs will improve stent-graft fixation in the abdominal aorta.Sixteen- to 24-mm-diameter Dacron grafts were deployed inside cadaveric aortas. The grafts were anchored by stents as in endovascular abdominal aortic aneurysm repair. One hundred thirty-seven stent-graft deployments were carried out with modified self-expanding Z-stents with (A) no hooks and barbs (n = 75), (B) 4 5-mm-long hooks and barbs (n = 39), (C) 8 10-mm-long, strengthened hooks and barbs (n = 19), or (D) hooks only (n = 4). Increasing longitudinal traction was applied to determine the displacement force needed to extract the stent-grafts. The radial force of the stents was measured and correlated to the displacement force.The median (interquartile range) displacement force needed to extract grafts anchored by stent A was 2.5 N (2.0 to 3.4), stent B 7.8 N (7.4 to 10.8), and stent C 22.5 N (17.1 to 27.9), p < 0.001. Both hooks and barbs added anchoring strength. During traction, the weaker barbs were distorted or caused intimal tears. The stronger barbs engaged the entire aortic wall. The radial force of the stents had no impact on fixation, while aortic calcification and graft oversizing had marginal effects.Stent barbs and hooks increased the fixation of stent-grafts tenfold, while the radial force of stents had no impact. These data may prove important in future endograft development to prevent stent-graft migration after aneurysm exclusion.