化学
再现性
免疫分析
电位滴定法
线性范围
碳二亚胺
检出限
微电极
电极
肝细胞癌
色谱法
抗体
生物化学
医学
癌症研究
物理化学
免疫学
作者
Ling Li,Liang Dong,Wenping Guo,Dianping Tang,Yongyi Zeng
标识
DOI:10.1002/elan.202100213
摘要
Abstract Herein, we investigated the analytical features of potentiometric immunosensors for detection of alpha‐fetoprotein (AFP) in hepatocellular carcinoma at different electrodes, such as carbon fiber microelectrode (CFME) and carbon‐disk electrode (CDE), respectively. To construct such an immunosensor, anti‐AFP capture antibodies were first conjugated covalently onto the activated electrodes through typical carbodiimide coupling. Thereafter, one‐step immunoreaction protocol was successfully introduced to develop a new potentiometric immunoassay upon addition of AFP. Accompanying the antigen‐antibody reaction, the surface charges of the modified electrodes were changed for the readout of electric potential. Results indicated that the linear range of CDE‐based immunosensor was 0.1–100 ng mL −1 AFP, whereas the assay sensitivity by using CFME could be further increased to 3.2 pg mL −1 with the linear range from 0.01 to 500 ng mL −1 AFP. Meanwhile, CFME‐based immunosensor showed high sensitivity, good reproducibility and specificity, and could be utilized for the analysis of human serum specimens with consistent results relative to commercialized ELISA kit.
科研通智能强力驱动
Strongly Powered by AbleSci AI