生物分析
他克莫司
免疫抑制
分析物
治疗药物监测
色谱法
霉酚酸
全血
萃取(化学)
医学
移植
化学
药理学
药代动力学
外科
内科学
作者
Abbie D. Leino,John Takyi-Williams,Bo Wen,Duxin Sun,Manjunath P. Pai
出处
期刊:Bioanalysis
[Newlands Press Ltd]
日期:2022-10-08
被引量:1
标识
DOI:10.4155/bio-2022-0155
摘要
Background: Volumetric absorptive microsampling may reduce the blood collection burden associated with therapeutic drug monitoring of immunosuppression to prevent organ transplant rejection. This work describes the development of a laboratory and analytical technique for quantifying tacrolimus and mycophenolic acid (MPA) from the Tasso-M20™ in human whole blood using bead-based impact-assisted extraction. Results: The sampled blood volume was accurate with estimated volumes within <2% of the expected 20 μl. Recovery using impact-assisted extraction was 73–87% for MPA and 100% for tacrolimus and was hematocrit-independent for both analytes. The LC-MS/MS assay is precise and accurate within the acceptance criteria of 15%. Conclusion: The sampling and extraction procedures allowed for accurate quantification of tacrolimus and MPA. Exploration of abuse scenarios identified important education points for patients conducting home-based sample collections in the future.