医学
连续血糖监测
血液透析
糖尿病
阿尔法(金融)
重症监护医学
1型糖尿病
急诊医学
内科学
内分泌学
外科
患者满意度
结构效度
作者
Parizad Avari,Reshaba Pushparatnam,Lalantha Leelarathna,Tricia Tan,Andrew Frankel,Nick Oliver,Monika Reddy
标识
DOI:10.1089/dia.2024.0575
摘要
The accuracy of the latest generation Dexcom G7 sensors in individuals with diabetes undergoing hemodialysis has not previously been investigated. Participants with diabetes undergoing hemodialysis were recruited, with paired sensor glucose from Dexcom G7 recorded with plasma glucose analyzed in the laboratory, as well as the Freestyle Precision Pro glucometer and EKF Biosen C-Line analyzer. Ten adults (median age 64.0 [58.0-74.5] years) were recruited. Overall percentage (%) mean and median absolute relative differences were 10.4% and 8.5% for matched laboratory pairs, respectively (