检出限
分光计
标准物质
基质(化学分析)
分析化学(期刊)
线性
航程(航空)
分辨率(逻辑)
质谱法
基础(拓扑)
材料科学
化学
色谱法
计算机科学
物理
数学
光学
量子力学
数学分析
人工智能
复合材料
作者
Alex Krummenauer,L H Lapazin,Marcelo Pilotto Cenci,Hugo Marcelo Veit
出处
期刊:Journal of physics
[IOP Publishing]
日期:2023-10-01
卷期号:2606 (1): 012001-012001
被引量:2
标识
DOI:10.1088/1742-6596/2606/1/012001
摘要
Abstract Utilizing a portable energy dispersive X-ray fluorescence (EDXRF) spectrometer and certified reference materials, this study aimed to determine the linearity, limit of detection (LOD), limit of quantification (LOQ), and working range of the method for quantitative analysis of base and precious metals in liquid matrices. As the EDXRF method lacks standardization and is highly susceptible to matrix effects, these calculations play a crucial role in the analytical validation process. The obtained R 2 values in this study exceeded 0.99, and the residuals exhibited a random distribution pattern. The LOD and LOQ values were determined to be exceptionally low, approaching the resolution limit of the spectrometer. Consequently, our calculations confirmed the analytical validation of the portable EDXRF method for analyzing elements in liquid matrices. The utilization of portable EDXRF equipment in quantitative measurements of elements in liquid matrices represents an innovative approach, unveiling new areas of application for this type of instrument.
科研通智能强力驱动
Strongly Powered by AbleSci AI