Jakub Kwiecinski,Raman Uberoi,Mohammed Hadi,Christopher P. Cheng,Zhong You
标识
DOI:10.1177/17085381241260910
摘要
Objective To quantify dynamic gutter phenomena and endograft deformations during double chimney thoracic endovascular aortic repair (ch-TEVAR) in a physiological model of the thoracic aorta subjected to pulsatile haemodynamic conditions. Methods Two in vitro procedures revascularizing the brachiocephalic trunk and left common carotid artery were performed representing both balloon-expandable (BE, Ankura-BeGraft) and self-expandable (SE, Ankura-Viabahn) double ch-TEVAR configurations. Retrospectively gated computed tomography (CT) was used to evaluate endograft behaviour. Device interactions were characterised according to gutter volume, gutter surface deviation, and endograft deformation (D-ratio) at end-diastolic and peak-systolic aortic pressure. Results Use of BE chimney grafts resulted in three times total gutter volume compared to SE chimney grafts. Gutter volumes were observed to vary dynamically between the end-diastolic and peak-systolic phases of the cardiac cycle, with the most substantial change associated with the BE configuration. Chimney graft deformations were dependent on device type, with SE devices exhibiting up to twice the deformation as BE devices. When adjacent, SE chimney grafts were observed to support each other, and thus tended towards a more consistently circular shape. Conclusion Gutter and chimney graft behaviour were dependent on device type, and exhibited both spatial and temporal variability. This study emphasises notable differences between BE and SE double ch-TEVAR configurations which should be considered when evaluating risk of endoleak. The findings reported here also support the use of gated CT to better identify device-related complications with ch-TEVAR, and can be used in the design of next generation devices.
科研通智能强力驱动
Strongly Powered by AbleSci AI