作者
Lu Zeng,Lingling Guo,Zhongxing Wang,Xinxin Xu,Shanshan Song,Liguang Xu,Hua Kuang,Aike Li,Chuanlai Xu
出处
期刊:Food Chemistry
[Elsevier BV]
日期:2021-01-18
卷期号:348: 129114-129114
被引量:55
标识
DOI:10.1016/j.foodchem.2021.129114
摘要
Pharmaceutical and food products are fortified with pantothenic acid (PA) to address potential deficiency. Therefore, its fast, reliable, and accurate detection is of great importance to the quality control. Here, an indirect competitive enzyme-linked immunosorbent assay (ic-ELISA) and a gold nanoparticle-based lateral flow immunoassay (LFIA) were established for the determination of PA based on an anti-PA monoclonal antibody (mAb). The ic-ELISA displayed a limit of detection (LOD) of 32.22 ng/mL, and the linear range was 64.44–628.84 ng/mL. Average recoveries of PA in fortified samples were 88.60–110.11% when using the ic-ELISA and a good correlation between the ic-ELISA and LC-MS/MS was obtained when analyzing samples. Furthermore, the developed LFIA strip showed a calculated LOD of 71.99, 115.80, and 240.12 ng/mL in B-complex Vitamin tablets, energy drink and infant milk powder samples, respectively. All the results demonstrated that both of these immunoassays are suitable for determining PA in pharmaceutical and food products.