A practical analysis of scatter-to-primary ratio after beam hardening correction for x-ray CT
材料科学
X射线
光学
核医学
梁(结构)
辐射
物理
成像体模
作者
Hewei Gao,Shumei Jia,Geng Fu
出处
期刊:Medical Imaging 2019: Physics of Medical Imaging日期:2019-03-01卷期号:10948: 1094832-被引量:1
标识
DOI:10.1117/12.2513266
摘要
Scatter and beam hardening degrade image quality for X-ray computed tomography (CT). Applying beam hardening correction before scatter being fully removed may be needed in some scenarios. In this paper, we theoretically analyze the outcome of beam hardening correction with scatter uncorrected. Using the effective spectrum of CT system, beam hardening correction function and its derivatives can be explicitly derived and used, along with a Taylor expansion, to approximate the change of scatter-to-primary ratio (SPR) after beam hardening correction. Two types of SPR compensations are compared using cone- and fan-beam scans of a CatPhan phantom on a tabletop cone-beam CT system, validating our analysis to be accurate in practical applications.