溶解
裂解缓冲液
流式细胞术
试剂
红细胞
色谱法
化学
细胞仪
外周血
全血
分子生物学
生物
免疫学
生物化学
物理化学
作者
Wei Cui,Sa Liu,Lun Cai,Yulin Li,Congcong Zhang,Siqi Qiu
出处
期刊:PubMed
日期:2011-04-01
卷期号:19 (2): 491-5
摘要
This study was purposed to investigate the efficacy of different whole flow lysis reagents for lysis of red blood cells in flow cytometric analysis. The expression of immunocytes was detected by flow cytometry after lysis of red blood cells using commercial reagents (Optilyse C, FACS Lysing Solution) and self-made red blood cell lysis reagents (RBC Lysis Buffer), the detection results were analyzed comparatively. The results showed that there was no significant difference in the percentage of CD3e(+), CD3e(+)CD4(+), CD3e(+)CD8a(+), CD3e(-)CD19(+), CD3e(-)NK1.1(+) and Gr-1(+) cells between 3 different lysis reagent groups. However OptiLyse C solution was suitable to Gr-1(+) cell detection, but did not suit to Foxp3(+) Treg detection. The self-made RBC Lysis Buffer and FACS Lysing Solution were suited to Foxp3(+) Treg detection. It is concluded that the use of self-made RBC Lysis Buffer for flow cytometry can get the lysis efficiency of commercially available lysis solutions when samples are prepared in accordance with standardized procedure. The self-made RBC Lysis Buffer not only can satisfy experimental requirements, but also can reduce the experimental costs.