Melissa McBride,Amitava Dasgupta
出处
期刊:PubMed
日期:2023-05-01
卷期号:53 (3): 482-484
摘要
We reported significant interference of biotin in FT3 and FT4 assays using Beckman DXI 800 analyzer. Recently we acquired Alinity i analyzer where TSH, FT3 and FT4 assays are not biotin based. We hypothesized that if thyroid function tests on DXI 800 and Alinity i are harmonized, then it is possible to eliminate biotin interference.We investigated assay harmonization by analyzing 35 specimens for TSH, FT4 and FT3 using both analyzers. We prepared one serum pool using left-over specimens where thyroid tests were ordered. Then aliquots of the pool were supplemented with various amounts of biotin followed by measuring thyroid function tests again.We observed assay harmonization between both analyzers for TSH, FT3 and FT4 Tests. TSH assay was not affected in the presence of biotin, but FT3 and FT4 values were significantly elevated using DXI 800 analyzer. In contrast, TSH, FT3 and FT4 assays were not affected by biotin using Alinity i analyzer.Elevated FT3 and FT4 using DXI 800 analyzer may be due to biotin interference which can be eliminated by observing normal values using Alinity i analyzer. However, normal or slightly elevated TSH with elevated FT3 and FT4 using both analyzers may indicate rare type of TSH producing tumor of pituitary, not biotin interference.