隐形眼镜
眼压
眼科
镜头(地质)
材料科学
光学
验光服务
医学
物理
作者
Hongbin An,Li Wang,Zhikang Liao,Liang Zhang,Hui Zhao,Yingyun Yang,Jizhou Song,Yinji Ma
标识
DOI:10.1038/s41528-024-00341-z
摘要
Twenty-four hours continuous intraocular pressure (IOP) monitoring is beneficial for glaucoma care. Contact lens sensors using LC technology can achieve non-invasive continuous IOP measuring around the clock and are relatively simple in principle and structure, thus dominating the mainstream. Among them, stretchable inductive LC contact lens sensors constructed with liquid metal have advantages in signal quality as well as wearing comfort, but currently its sensitivity is slightly insufficient. Here, we propose an LC sensor that constructed with liquid metal to form the stretchable inductance and capacitance further. The capacitive plate and inductive coil can response to IOP changes simultaneously, thus able to enhance the sensitivity in principle. We modeled the sensing mechanism and conducted design, fabrication, as well as various tests. The device exhibits good characteristics, including reliability, good signal quality, etc. Especially, it has a threefold increase in sensitivity, exceeding the current state-of-the-art contact lens sensors.