作者
Arturo García‐Touchard,Javier Goicolea,Manel Sabaté,Fernándo Alfonso,Rafael Ruíz-Salmerón,Armando Bethencourt,Nieves Gonzalo,Faustino Miranda,Bruno García del Blanco,Jesús Jiménez-Mazuecos,Rafael Melgares,Pedro Martínez-Romero,José M. Hernández-García,R. Lezáun,Juan A. Bullones,Javier Fernández Portales,José Ramón Rumoroso,Rosario Ortas,Mariano Valdés,Ramiro Trillo-Nouche,Salvatore Brugaletta,Shuji Otzuki,Francisco José Hernández Pérez,Luis Alonso‐Pulpón
摘要
Our aim was to assess the safety and efficacy of paclitaxel-eluting balloon (PTX-B) treatment after bare metal stent (BMS) implantation in patients undergoing primary angioplasty.After BMS implantation, patients were randomised (1:1) to treatment with a PTX-B or no PTX-B treatment (BMS group). The primary endpoint was in-stent late luminal loss (LLL) at nine-month follow-up. OCT was carried out on the first 20% of consecutive patients included in the study. Two hundred and twenty-three patients were randomised (BMS: 112, PTX-B: 111). At nine months, median LLL was 0.80 mm (interquartile range [IQR] 0.36-1.26) in the BMS group vs. 0.31 mm (IQR 0.00-0.58) in the PTX-B group, p<0.0001. Binary restenosis was significantly lower in the PTX-B group: 29.8% vs. 2.2%, p<0.0001, 95% confidence interval (CI): 3.2-54.2. Nine-month OCT showed good strut coverage in both groups but greater in the BMS group (100±0.0% vs. 99.52±1.11%, p=0.03) with very low rates of malapposed struts per lesion. One-year MACE was significantly lower in the PTX-B group (12.5% vs. 3.6%, p=0.016).PTX-B after successful BMS implantation resulted in less LLL and better clinical outcomes as compared with a BMS-only strategy. This was associated with good stent strut coverage and very low rates of malapposed struts.