A method for the removal of serum chylomicrons before density gradient ultracentrifugation of the other serum lipoproteins using an SW 41 swinging bucket rotor is presented. In a preliminary spin, the chylomicrons with an Sf > 400·10−13s float to the top of the gradient, whereas the other lipoproteins are retained in the infranatant fraction. After removal of the chylomicrons, the other serum lipoproteins are subsequently fractionated by isopycnic density gradient ultracentrifugation. Analysis of the separated lipoprotein fractions suggested that this procedure premits isolation of a chylomicron fraction consisting solely of chylomicrons but that the very low density lipoprotein fraction subsequently isolated also contains chylomicrons or chylomicron remnants with an Sf < 400·10−13s, and that there is considerable overlap in floatation rate and particle sizeof very low density lipoproteins and chylomicrons.