Mengjiao Li,Yi Wu,Chun‐Yen Ke,Zichen Song,Meie Zheng,Qingjie Yu,Hongda Zhu,Huiling Guo,Hongmei Sun,Mingxing Liu
出处
期刊:Talanta
[Elsevier]
日期:2024-04-01
卷期号:270: 125546-125546
被引量:4
标识
DOI:10.1016/j.talanta.2023.125546
摘要
In this study, an ultrasensitive unlabeled electrochemical immunosensor for the detection of cardiac troponin I (cTnI) was developed based on Pt/Au modified B,S,N co-doped reduced graphene oxide (Pt/Au-B,S,N-rGO) as a signal amplification platform. First-principles calculations were employed to analyze the electron density of states of Pt/Au-B,S,N-rGO, revealing an increase in the electron density of the graphene oxide (GO) states. Furthermore, scanning electron microscopy (SEM), X-ray photoelectron diffraction spectroscopy (XPS), and electrochemical detection were used to successfully construct and analyze Pt/Au-B,S,N-rGO. The results showed that B,S,N-rGO exhibited good electrochemical activity, and the Au/Pt NPs demonstrated excellent catalytic properties, which provided a strong foundation for achieving high-sensitivity detection. Moreover, the constructed unlabeled electrochemical immunosensor had an ideal linear range (0.1 pg/mL∼50 ng/mL) and detection limit (0.082 pg/mL). In human serum detection, the results of this immunosensor were essentially similar to the ELISA results for the same samples, which suggested that the immunosensor had a promising clinical application prospect for the detection of cTnI.