医学
动脉瘤
支架
血流
软件部署
外科
放射科
计算机科学
操作系统
作者
Haoyu Zhu,Yupeng Zhang,Shikai Liang,Chuhan Jiang
标识
DOI:10.1136/jnis-2024-022430
摘要
The Woven EndoBridge (WEB) device is a well established treatment for bifurcation aneurysms.1–6 However, failed detachment after deployment can present significant challenges. In this technical video (video 1), we report on a patient with a left middle cerebral artery (MCA) bifurcation aneurysm treated with the WEB device. Despite satisfactory deployment, multiple detachment attempts were unsuccessful. After repeated maneuvers, the WEB was finally detached but slightly protruded from the aneurysm sac, compromising blood flow in the superior branch of the MCA. Even after placing an Atlas stent, blood flow was not restored. Ultimately, using a microguidewire and microcatheter, we repositioned the protruded WEB device back into the aneurysm sac, successfully restoring blood flow. This case illustrates that the Atlas stent provides limited support for the WEB device. In similar situations, gently repositioning the protruded WEB back into the aneurysm sac may be a remedial measure. Video 1 Technical video demonstrating rescue techniques for managing intravascular mechanical obstruction following detachment of the WEB device
科研通智能强力驱动
Strongly Powered by AbleSci AI