电化学发光
癌胚抗原
钌
检出限
电极
免疫分析
材料科学
炸薯条
化学
纳米技术
组合化学
色谱法
计算机科学
癌症
催化作用
医学
抗体
生物化学
电信
物理化学
内科学
免疫学
作者
Wenjing Luo,Zhuoxin Ye,Pinyi Ma,Qiong Wu,Daqian Song
出处
期刊:Analyst
[The Royal Society of Chemistry]
日期:2022-01-01
卷期号:147 (9): 1986-1994
被引量:11
摘要
Carcinoembryonic antigen (CEA) is an important cancer marker that plays a significant role in achieving low-cost, rapid and highly sensitive clinical detection. In this work, we developed a disposable electrochemiluminescence (ECL) sensor chip based on a screen-printed electrode (SPE) for detecting CEA via ECL technology. An amino-modified Ti3C2 MXene was used as a carrier to successfully prepare a highly efficient ECL probe AuNPs-Ru-Arg@NH2-Ti3C2-MXene by loading with AuNPs-Arg through covalent links and modifying with a ruthenium complex. Upon the addition of CEA, the ECL signal decreased significantly with the increase of CEA, due to the formation of immune complexes at the interface of the electrode. The sensing chip was used to detect CEA in an aqueous solution and found to have a detection limit of 1.5 pg mL-1. The chip was used to determine CEA in the serum of healthy humans and cancer patients, and the results were consistent with those obtained using ELISA. The disposable ECL sensor chip has many advantages including convenience, rapid detection, low cost and easy mass production; thus it has great application potential in clinical cancer diagnosis.
科研通智能强力驱动
Strongly Powered by AbleSci AI