医学
假体周围
植入
固定(群体遗传学)
全髋关节置换术
假肢
外科
关节置换术
牙科
环境卫生
人口
作者
Christopher C Cooke,William J. Hozack,Carlos J. Lavernia,Peter F. Sharkey,Shani Shastri,Richard H. Rothman
标识
DOI:10.1016/s0883-5403(03)00325-5
摘要
Fifty-eight patients received an Osteonics constrained acetabular implant for recurrent instability (46), girdlestone reimplant (8), correction of leg lengthening (3), and periprosthetic fracture (1). The constrained liner was inserted into a cementless shell (49), cemented into a pre-existing cementless shell (6), cemented into a cage (2), and cemented directly into the acetabular bone (1). Eight patients (13.8%) required reoperation for failure of the constrained implant. Type I failure (bone-prosthesis interface) occurred in 3 cases. Two cementless shells became loose, and in 1 patient, the constrained liner was cemented into an acetabular cage, which then failed by pivoting laterally about the superior fixation screws. Type II failure (liner locking mechanism) occurred in 2 cases. Type III failure (femoral head locking mechanism) occurred in 3 patients. Seven of the 8 failures occurred in patients with recurrent instability. Constrained liners are an effective method for treatment during revision total hip arthroplasty but should be used in select cases only.
科研通智能强力驱动
Strongly Powered by AbleSci AI