影像引导放射治疗
放射治疗
医学
医学物理学
核医学
计算机科学
放射科
作者
R. de Crevoisier,C. Lafond,A. Mervoyer,Corinne Hulot,N. Jaksic,Igor Bessières,G. Delpon
标识
DOI:10.1016/j.canrad.2021.08.002
摘要
We present the updated recommendations of the French society for oncological radiotherapy on image-guided radiotherapy (IGRT). The objective of the IGRT is to take into account the anatomical variations of the target volume occurring between or during the irradiation fractions, such as displacements and/or deformations, so that the delivered dose corresponds to the planned dose. This article presents the different IGRT devices, their use and quality control, and quantify the possible additional dose generated by each of them. The practical implementation of IGRT in various tumour locations is summarised, from the different "RecoRad™" guideline articles. Adaptive radiotherapy is then detailed, due to its complexity and its probable development in the next years. The place of radiation technologist in the practice of IGRT is then specified. Finally, a brief update is proposed on the delicate question of the additional dose linked to the in-room imaging, which must be estimated and documented at a minimum, as long as it is difficult to integrate it into the calculation of the dose distribution.
科研通智能强力驱动
Strongly Powered by AbleSci AI